Speed Calibration Your Combo V24T was pre-calibrated before it was shipped from Jorgensen Laboratories. If future calibration is needed, the following procedure requires a hand-held tachometer and should only be performed by a qualified technician or the Jorgensen Laboratories Service & Repair department. The bottom plate must be removed to access the calibration rheaostats. With the Rotor Select on the front of the unit on high, the speed to 12,000 rpm, and a Microhematocrit rotor in place, adjust the Vr-h rheostat on the circuit board up to 12000 RPMs. With the Rotor Select on the front of the unit on low, the speed set to 4,000 rpm, and a tube rotor with tube shields in place, adjust the VR-I rheostat on the circuit board up to 4,000 rpm. 1 2 Brush Replacement Unplug the centrifuge, and turn it on its side. Remove the feet and all Phillips head screws holding the metal base plate in place. Using a flat head screwdriver, remove the two black caps on the bottom part of the motor. Using your fingers, pull the brush out of the recessed hole that the black caps were removed from. The brushes consist of springs with a graphite bar on them. Place the new brushes in the recessed hole, and replace the black caps. Be careful not to overtighten. Black caps could break. (See image). Replace the metal plate and the feet. Your Combo V24T is now ready to use. 1 2 3 4 5 6 7 Specifications Max. Speed (HIGH): Max. Speed (LOW): Max. Volume: Input Voltage: Fuse: Height: Width: Depth: Weight: 12,000 RPM (+/- 600) 4,000 RPM (+/- 200) 24 3ml to 15ml test tubes 24 75mm capillary tubes 24 2ml microtubes 110v (220v available) 6.3 amp / 250 volt 8.7 inches 10.6 inches 12.5 inches 15.8 lbs. WARNING: Failure to secure rotor or rotor lid could lead to personal injury or damage to the centrifuge. CAUTION: If corrosion, scratches, or other abnormalities are found on rotor or tube shields, discontinue use and contact our service center. MKT-7.5.3-L-239.001 | Rev 0 Recommended Fluid Speeds & Times: Blood (Serum) 4,000 rpms 7 to 10 minutes Urine 1,700 rpms 5 to 10 minutes Fecals 1,500 rpms 6 minutes Microhematocrits 12,000 rpms 3 minutes RPM 500 1,000 1,400 1,700 2,000 2,500 3,000 3,500 4,000 12,000 12,000 G-Force (RCF) 38 153 300 443 613 957 1,378 1,876 2,451 12,9000 14,500 Test Tube Rotor Microtube Rotor Microhematocrit Rotor P. 800.525.5614 F. 970.663.5042 1450 Van Buren Ave Loveland Co 80538 USA